Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 85: (1 year) CAL VER /5520 DATE: 20-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 72 NUMBER OF LINES: 248 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Fluke 80 Series Service Manual. # (PN 834168 April 1989, Rev. 3, 7/93, Change/Errata 8/94). # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F W 1.002 ASK+ K 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 RSLT =DISPLAY TEST 1.006 PICE 83_85dsp 2.001 DISP No warm-up time requirement. 2.001 DISP Ambient temperature: 18C - 28C. 2.001 DISP Relative humidity: less than 90%. 2.002 ASK- U 2.003 HEAD {ROTARY SWITCH TESTS} 2.004 DISP Hold down the RANGE button while turning the UUT rotary 2.004 DISP function switch from OFF to VDC. Release the RANGE 2.004 DISP button and turn the switch to {VAC}. 2.005 MATH MEM1 = 0 2.006 MEMI Enter UUT reading: 2.007 MEME 2.008 MEMC display 12U 3.001 DISP Rotate the UUT function switch to {VDC}. 3.002 MATH MEM1 = -32 3.003 MEMI Enter UUT reading: 3.004 MEME 3.005 MEMC display 12U 4.001 DISP Rotate the UUT function switch to {mVDC}. 4.002 MATH MEM1 = -64 4.003 MEMI Enter UUT reading: 4.004 MEME 4.005 MEMC display 12U 5.001 DISP Rotate the UUT function switch to {Ohms}. 5.002 MATH MEM1 = -96 5.003 MEMI Enter UUT reading: 5.004 MEME 5.005 MEMC display 12U 6.001 DISP Rotate the UUT function switch to {Diode}. 6.002 MATH MEM1 = -128 6.003 MEMI Enter UUT reading: 6.004 MEME 6.005 MEMC display 12U 7.001 DISP Rotate the UUT function switch to {mA/A}. 7.002 MATH MEM1 = -160 7.003 MEMI Enter UUT reading: 7.004 MEME 7.005 MEMC display 12U 8.001 DISP Rotate the UUT function switch to {uA}. 8.002 MATH MEM1 = -192 8.003 MEMI Enter UUT reading: 8.004 MEME 8.005 MEMC display 12U 9.001 ASK+ U 9.002 HEAD {AC VOLTAGE TESTS} 9.003 DISP Rotate the UUT function switch to OFF then VAC. 9.004 DISP Connect the 5520A to the UUT as follows: 9.004 DISP [32] 5520A NORMAL HI to UUT V/Ohm/Diode 9.004 DISP [32] 5520A NORMAL LO to UUT COM 9.005 HEAD AC VOLTAGE TESTS: {400mV Range} 9.006 5520 400 350.0mV 2.2U 60H SI 2W 10.001 5520 400 350.0mV 3.9U 1kH SI 2W 11.001 5520 400 350.0mV 7.4U 5kH SI 2W 12.001 5520 400 350.0mV 10.9U 20kH SI 2W 13.001 HEAD AC VOLTAGE TESTS: {4V Range} 13.002 5520 4 3.500V 0.020U 60H SI 2W 14.001 5520 4 3.500V 0.037U 1kH SI 2W 15.001 5520 4 3.500V 0.072U 5kH SI 2W 16.001 5520 4 3.500V 0.144U 20kH SI 2W 17.001 HEAD AC VOLTAGE TESTS: {40V Range} 17.002 5520 40 35.00V 0.20U 60H SI 2W 18.001 5520 40 35.00V 0.37U 1kH SI 2W 19.001 5520 40 35.00V 0.72U 5kH SI 2W 20.001 5520 40 35.00V 1.44U 20kH SI 2W 21.001 HEAD AC VOLTAGE TESTS: {400V Range} 21.002 5520 400 350.0V 2.0U 60H SI 2W 22.001 5520 400 350.0V 3.7U 1kH SI 2W 23.001 5520 400 350.0V 7.2U 5kH SI 2W 24.001 5520 400 100.0V 4.4U 20kH SI 2W 25.001 5520 400 200.0V 8.4U 20kH SI 2W 26.001 5520 400 300.0V 12.4U 20kH SI 2W # 350V test at 10kHz instead of 20kHz. 27.001 5520 400 350.0V 14.4U 10kH SI 2W 28.001 HEAD AC VOLTAGE TESTS: {1000V Range} 28.002 5520 1000 900V 7U 60H SI 2W 29.001 5520 1000 900V 20U 1kH SI 2W 30.001 5520 1000 900V 20U 5kH SI 2W 31.001 HEAD {FREQUENCY TESTS} 31.002 DISP Select the 400mV Range on the UUT. 31.002 DISP Press the Hz button to select the Frequency Counter. 31.003 HEAD FREQUENCY TESTS: {19.999kHz Range} 31.004 5520 20 19.000kH 0.002U 150mV SI 2W 32.001 HEAD FREQUENCY TESTS: {199.99kHz Range} 32.002 5520 200 190.00kH 0.02U 150mV SI 2W 33.001 HEAD {FREQUENCY SENSITIVITY AND TRIGGER LEVEL} 33.002 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {4V AC Range} 33.003 DISP Press the RANGE button to lock the 4V Range. 33.004 5520 1000.0H 0.2U 300mV SI 2W 34.001 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {4V DC Range} 34.002 DISP Rotate the UUT function switch to VDC. 34.002 DISP Press the RANGE button to lock the 4V Range. 34.002 DISP Press the Hz button to choose the Frequency mode. 34.003 5520 1000.0H 0.2U 1.7V SI 2W 35.001 5520 1000.0H 1.0V SI S 2W 35.002 EVAL Is the UUT {displaying all zeros}? 36.001 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {40V DC Range} 36.002 DISP Press the RANGE button to lock the 40V Range. 36.003 5520 1000.0H 0.2U 6.0V SI 2W 37.001 5520 1000.0H 2.0V SI S 2W 37.002 EVAL Is the UUT {displaying all zeros}? 38.001 HEAD {DC VOLTAGE TESTS} 38.002 DISP Press the Hz button twice for VDC 38.002 DISP 38.002 DISP Press RANGE for 2s to select autorange. 38.003 HEAD DC VOLTAGE TESTS: {4V Range} 38.004 5520 4 3.500V 0.005U 2W 39.001 HEAD DC VOLTAGE TESTS: {40V Range} 39.002 5520 40 35.00V 0.05U 2W 40.001 5520 40 -35.00V 0.05U 2W 41.001 HEAD DC VOLTAGE TESTS: {400V Range} 41.002 5520 400 350.0V 0.5U 2W 42.001 HEAD DC VOLTAGE TESTS: {1000V Range} 42.002 5520 1000 1000V 2U 2W 43.001 HEAD {DC MILLIVOLT TEST} 43.002 DISP Rotate the UUT function switch to mVDC. 43.003 HEAD DC MILLIVOLT TEST: {400mV Range} 43.004 5520 400 350.0mV 0.5U 2W 44.001 HEAD {RESISTANCE TESTS} 44.002 DISP Rotate the UUT function switch to Ohms. 44.003 DISP Connect the 5520A to the UUT as follows: 44.003 DISP [32] 5520A NORMAL HI to UUT V/Ohm/Diode 44.003 DISP [32] 5520A NORMAL LO to UUT COM 44.003 DISP [32] 5520A AUX HI to UUT V/Ohm/Diode 44.003 DISP [32] 5520A AUX LO to UUT COM 44.004 HEAD RESISTANCE TESTS: {400 Ohm Range} 44.005 5520 400 190.0Z 0.5U CW 45.001 HEAD RESISTANCE TESTS: {40 kOhm Range} 45.002 5520 40 19.00kZ 0.05U CW 46.001 HEAD RESISTANCE TESTS: {4 MOhm Range} 46.002 DISP Connect the 5520A to the UUT as follows: 46.002 DISP [32] 5520A NORMAL HI to UUT V/Ohm/Diode 46.002 DISP [32] 5520A NORMAL LO to UUT COM 46.003 5520 4 1.900MZ 0.005U 2W 47.001 HEAD RESISTANCE TESTS: {40 MOhm Range} 47.002 5520 40 19.00MZ 0.22U 2W 48.001 HEAD {CONDUCTANCE TEST} 48.002 DISP Press the Range button twice for 40nS range. 48.003 HEAD CONDUCTANCE TEST: {40nS Range} 48.004 5520 40 10.00nY 0.20U 2W 49.001 HEAD {CAPACITANCE TESTS} 49.002 DISP Press the BLUE button to select Capacitance. 49.003 HEAD CAPACITANCE TESTS: {5.00uF Range} 49.004 5520 5 1.00uF 0.03U CW 50.001 HEAD CAPACITANCE TESTS: {0.500uF Range} 50.002 5520 0.5 0.470uF 0.007U CW 51.001 HEAD CAPACITANCE TESTS: {0.0500uF Range} 51.002 5520 0.05 0.0470uF 0.0007U 2W 52.001 HEAD CAPACITANCE TESTS: {5.00nF Range} 52.002 DISP Select Relative mode (REL) to zero the display and 52.002 DISP automatically subtract the residual meter and test lead 52.002 DISP capacitance. 52.003 5520 5 4.70nF 0.07U 2W 53.001 HEAD {DIODE TEST} 53.002 DISP Rotate the UUT function switch to Diode. 53.003 5520 3.000V 0.061U 2W 54.001 DISP Disconnect the 5520A from the UUT. 54.002 HEAD {DC MILLIAMP TESTS} 54.003 DISP Rotate the UUT function switch to mA/A. 54.004 DISP Connect the 5520A to the UUT as follows: 54.004 DISP [32] 5520A AUX HI to UUT mA/uA 54.004 DISP [32] 5520A AUX LO to UUT COM 54.005 HEAD DC MILLIAMP TESTS: {40mA Range} 54.006 5520 40 35.00mA 0.09U 2W 55.001 HEAD DC MILLIAMP TESTS: {400mA Range} 55.002 5520 400 350.0mA 0.9U 2W 56.001 HEAD {AC MILLIAMP TESTS} 56.002 DISP Press the BLUE button to select AC. 56.003 HEAD AC MILLIAMP TESTS: {40mA Range} 56.004 5520 40 35.00mA 0.23U 60H SI 2W 57.001 5520 40 35.00mA 0.23U 1kH SI 2W 58.001 HEAD AC MILLIAMP TESTS: {400mA Range} 58.002 5520 400 350.0mA 2.3U 60H SI 2W 59.001 5520 400 350.0mA 2.3U 1kH SI 2W 60.001 HEAD {DC MICROAMP TESTS} 60.002 DISP Rotate the UUT function switch to uA. 60.003 HEAD DC MICROAMP TESTS: {400uA Range} 60.004 5520 400 350.0uA 1.0U 2W 61.001 HEAD DC MICROAMP TESTS: {4000uA Range} 61.002 5520 4000 3500uA 9U 2W 62.001 HEAD {AC MICROAMP TESTS} 62.002 DISP Press the BLUE button to select AC. 62.003 HEAD AC MICROAMP TESTS: {400uA Range} 62.004 5520 400 350.0uA 2.3U 60H SI 2W 63.001 5520 400 350.0uA 2.3U 1kH SI 2W 64.001 HEAD AC MICROAMP TESTS: {4000uA Range} 64.002 5520 4000 3500uA 23U 60H SI 2W 65.001 5520 4000 3500uA 23U 1kH SI 2W 66.001 HEAD {DC AMP TESTS} 66.002 DISP Rotate the UUT function switch to mA/A. 66.003 DISP Connect the 5520A to the UUT as follows: 66.003 DISP [32] 5520A 20A to UUT A 66.003 DISP [32] 5520A AUX LO to UUT COM 66.004 HEAD DC AMP TESTS: {4000mA Range} 66.005 5520 4000 3500mA 9U HC 2W 67.001 HEAD DC AMP TESTS: {10A Range} 67.002 5520 10 10.00A 0.04U HC 2W 68.001 HEAD {AC AMP TESTS} 68.002 DISP Press the BLUE button to select AC. 68.003 HEAD AC AMP TESTS: {4000mA Range} 68.004 5520 4000 3500mA 23U 60H SI HC 2W 69.001 5520 4000 3500mA 23U 1kH SI HC 2W 70.001 HEAD AC AMP TESTS: {10A Range} 70.002 5520 10 10.00A 0.08U 60H SI HC 2W 71.001 5520 10 10.00A 0.08U 1kH SI HC 2W 72.001 END